FDA Adverse Event Other Summary report: N

SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D

MDR report key: 2074622 · Received April 21, 2011

Report

Report Number
1220984-2011-00002
Event Type
Other
Date Received
April 21, 2011
Date of Event
March 25, 2011
Report Date
April 20, 2011
Manufacturer
HOLOGIC, INC.
Product Code
MUE
PMA / PMN Number
P010025
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2011 A HOLOGIC FIELD ENGINEER WENT TO THE SITE TO DETERMINE THE CAUSE OF THE INCIDENT. THE FIELD ENGINEER WAS UNABLE TO REPRODUCE THE REPORTED PROBLEM. THE FIELD ENGINEER DID NOTICE THE RIGHT SIDE LOCK WAS STICKING, BUT WAS LOCKING. HE ALSO NOTICED SOME A CHIPS IN THE PAINT ON THE X1.5 MAGNIFICATION POSITION AND FEELS THAT SOME OF THE TECHNOLOGISTS MAY BE USING THIS POSITION UNINTENTIONALLY. THE MAGNIFICATION STAND WAS OPERATING TO HOLOGIC'S SPECIFICATIONS WHEN SECURELY ATTACHED TO THE SYSTEM. THE FIELD ENGINEER REPLACED THE MAGNIFICATION STAND WITH A NEW ONE. PROPER FUNCTIONING WAS CONFIRMED PRIOR TO LEAVING THE FACILITY. THE MAGNIFICATION STAND IN QUESTION WAS RETURNED TO THE MFG FACILITY FOR EVAL. THE MAGNIFICATION STAND WAS REC'D AND EVALUATED BY A MFG ENGINEER. MINOR RUBBING WAS NOTED ON THE LINKAGE TO THE LOCKING MECHANISM. THE MFG ENGINEER ATTEMPTED SEVERAL TIMES, BUT COULD NOT DUPLICATE THE REPORTED PROBLEM. IT WAS CONFIRMED THAT WHEN THE MAGNIFICATION STAND IS PROPERLY SECURED, IT REMAINS ATTACHED TO THE SYSTEM WHEN MOVED TO A 90 DEGREE ANGLE AS DESIGNED. THE CLINICAL SERVICE DEPT IS COORDINATING A CUSTOMER CARE VISIT TO THE SITE. DURING THE PROPOSED VISIT, A HOLOGIC APPLICATIONS SPECIALIST WILL REVIEW USE OF THE SELENIA DIMENSIONS SYSTEM INCLUDING HOW TO PROPERLY SECURE THE MAGNIFICATION STAND TO THE SYSTEM.

Description of Event or Problem · 1

THE MAGNIFICATION STAND FELL OFF THE SELENIA DIMENSIONS SYSTEM WHEN THE TECHNOLOGIST ROTATED THE C-ARM TO A 90 DEGREE ANGLE. WHEN THE MAGNIFICATION STAND FELL, IT LANDED ON THE PT'S FOOT. THE PT SAID SHE WAS "OKAY" AND DECLINED MEDICAL ATTENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D MAMMOGRAPHY SYSTEM MUE HOLOGIC, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other