SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D
Report
- Report Number
- 1220984-2011-00002
- Event Type
- Other
- Date Received
- April 21, 2011
- Date of Event
- March 25, 2011
- Report Date
- April 20, 2011
- Manufacturer
- HOLOGIC, INC.
- Product Code
- MUE
- PMA / PMN Number
- P010025
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON (B)(4) 2011 A HOLOGIC FIELD ENGINEER WENT TO THE SITE TO DETERMINE THE CAUSE OF THE INCIDENT. THE FIELD ENGINEER WAS UNABLE TO REPRODUCE THE REPORTED PROBLEM. THE FIELD ENGINEER DID NOTICE THE RIGHT SIDE LOCK WAS STICKING, BUT WAS LOCKING. HE ALSO NOTICED SOME A CHIPS IN THE PAINT ON THE X1.5 MAGNIFICATION POSITION AND FEELS THAT SOME OF THE TECHNOLOGISTS MAY BE USING THIS POSITION UNINTENTIONALLY. THE MAGNIFICATION STAND WAS OPERATING TO HOLOGIC'S SPECIFICATIONS WHEN SECURELY ATTACHED TO THE SYSTEM. THE FIELD ENGINEER REPLACED THE MAGNIFICATION STAND WITH A NEW ONE. PROPER FUNCTIONING WAS CONFIRMED PRIOR TO LEAVING THE FACILITY. THE MAGNIFICATION STAND IN QUESTION WAS RETURNED TO THE MFG FACILITY FOR EVAL. THE MAGNIFICATION STAND WAS REC'D AND EVALUATED BY A MFG ENGINEER. MINOR RUBBING WAS NOTED ON THE LINKAGE TO THE LOCKING MECHANISM. THE MFG ENGINEER ATTEMPTED SEVERAL TIMES, BUT COULD NOT DUPLICATE THE REPORTED PROBLEM. IT WAS CONFIRMED THAT WHEN THE MAGNIFICATION STAND IS PROPERLY SECURED, IT REMAINS ATTACHED TO THE SYSTEM WHEN MOVED TO A 90 DEGREE ANGLE AS DESIGNED. THE CLINICAL SERVICE DEPT IS COORDINATING A CUSTOMER CARE VISIT TO THE SITE. DURING THE PROPOSED VISIT, A HOLOGIC APPLICATIONS SPECIALIST WILL REVIEW USE OF THE SELENIA DIMENSIONS SYSTEM INCLUDING HOW TO PROPERLY SECURE THE MAGNIFICATION STAND TO THE SYSTEM.
THE MAGNIFICATION STAND FELL OFF THE SELENIA DIMENSIONS SYSTEM WHEN THE TECHNOLOGIST ROTATED THE C-ARM TO A 90 DEGREE ANGLE. WHEN THE MAGNIFICATION STAND FELL, IT LANDED ON THE PT'S FOOT. THE PT SAID SHE WAS "OKAY" AND DECLINED MEDICAL ATTENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM 2D | MAMMOGRAPHY SYSTEM | MUE | HOLOGIC, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |