FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2074540
·
Received May 3, 2011
Report
- Report Number
- 2531779-2011-03044
- Event Type
- Malfunction
- Date Received
- May 3, 2011
- Date of Event
- February 6, 2011
- Report Date
- February 6, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MFR DATE: 07/2008. RECALL #: 2531779-03/24/2010-003-R. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.
Description of Event or Problem · 1
PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED PARTIALLY DISLODGED FORCE SENSOR PINS. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 |