FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2073696 · Received May 2, 2011

Report

Report Number
1423500-2011-05261
Event Type
Malfunction
Date Received
May 2, 2011
Date of Event
April 12, 2011
Report Date
April 12, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED ISSUE OF FAILED THE GROUND BOND TEST WAS CONFIRMED. THE ASSIGNABLE CAUSE OF THE RITE ELECTRICAL TEST FAILURE OF GROUND BOND FAILED PERFORMANCE SPEC WAS DETERMINED TO BE A LOOSE SETSCREW ON THE DOOR POST. THE DOOR POST WILL BE SCRAPPED DURING SERVICING. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THE GROUND BOND TEST INDICATING A HIGH RESISTANCE VALUE BETWEEN THE HC AND THE GROUND BOND ON THE POWER SUPPLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment