FDA Adverse Event Malfunction Summary report: N

GLENOSPHERE ORIENTATION GUIDE

MDR report key: 20736071 · Received November 20, 2024

Report

Report Number
1818910-2024-24187
Event Type
Malfunction
Date Received
November 20, 2024
Date of Event
November 11, 2024
Manufacturer
DEPUY IRELAND - 9616671
Product Code
PXH
UDI-DI
10603295116592
PMA / PMN Number
K170748
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: BROKEN INSTRUMENT ART. 230795000. THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE GLENOSPHERE ORIENTATION GUIDE HAS THE METAL PRONG BENT AND WAS CRACKED FROM THE TIP. BASED ON THE OBSERVED CONDITION, IT IS CONSISTENT WITH THE MATERIAL OVERLOAD THROUGH THE USE OF EXCESSIVE FORCES APPLIED, THUS THE POTENTIAL CAUSE COULD BE ATTRIBUTED TO UNINTENDED USE ERROR. A FUNCTIONAL TEST WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE GLENOSPHERE ORIENTATION GUIDE WOULD HAVE CONTRIBUTED TO THE COMPLAINED ISSUE. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE IS TRACED TO UNINTENDED USE ERROR, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. H11 ADDITIONAL NARRATIVE: CORRECTED: H3.

Additional Manufacturer Narrative · 0

DEPUY ORTHOPAEDICS IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY ORTHOPAEDICS HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY ORTHOPAEDICS OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY ORTHOPAEDICS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED ON NOVEMBER, 11, 2024 THE INSTRUMENT WAS BROKEN. SURGERY WAS DELAYED 10 MINUTES. NEW INSTRUMENT WAS USED. PROCEDURE WAS SUCCESSFULLY COMPLETED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. THE PLASTIC NEAR THE METALLIC NOSE ON THE INSTRUMENT IS BROKEN, IT IS STILL IN ONE PIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2138342 GLENOSPHERE ORIENTATION GUIDE EXTREMITY INSTRUMENTS : ALIGNMENT DEVICES PXH DEPUY IRELAND - 9616671 5416013 10603295116592

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown