FDA Adverse Event Summary report: N

SPECTRUM

MDR report key: 2073547 · Received April 18, 2011

Report

Report Number
2073547
Date Received
April 18, 2011
Date of Event
March 16, 2011
Report Date
April 12, 2011
Manufacturer
SIGMA
Product Code
FRN
Report Source
User Facility report
Reporter Location
WA, US

Narratives

Description of Event or Problem · 1

A POTASSIUM IV WAS HUNG TO BE ADMINISTERED BY A SIGMA SPECTRUM PUMP. SOON AFTER ADMINISTRATION, AN UPSTREAM OCCLUSION ALARM OCCURRED. THE NURSE RETURNED AND CHECKED THE LINES AND RESTARTED THE INFUSION. LATER IN THE SHIFT DURING A ROUTINE CHECK, IT WAS NOTED THAT EVEN THOUGH THE PUMP WAS STILL OPERATING, NO MEDICATION WAS BEING DELIVERED. ON INSPECTION THE NURSE NOTED A CLAMP ON THE UPSTREAM SIDE OF THE RUMP WAS STILL CLOSED. THE PUMP WAS REMOVED AND SENT TO BIOMED FOR TESTING. THE BIOMED STAFF CONFIRMED THAT A SECOND UPSTREAM OCCLUSION COULD GO WITHOUT ALARM IF PUMP WAS RESTARTED IN THE OCCLUDED STATE. SIGMA WAS CALLED AND THEY STATED THAT THIS IS NORMAL AND CORRECT OPERATION BASED ON THE TECHNOLOGY USED TO DETECT UPSTREAM OCCLUSIONS ON THESE PUMPS AND IS STATED AS SUCH BOTH IN TRAINING AND IN THE OPERATIONS MANUAL. THIS IS BEING REPORTED NOT BECAUSE THE PUMP MALFUNCTIONED, IN FACT IT PASSED ALL OPERATIONS TESTS, BUT BECAUSE THE STAFF FOUND IT COUNTER INTUITIVE AND POSSIBLY DANGEROUS THAT A DEVICE COULD BE INADVERTENTLY RESTARTED IN AN ALARM STATE AND NO LONGER RECOGNIZE THAT ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION PUMP FRN SIGMA SPECTRUM *

Patients

Seq Age Sex Outcome Treatment
1 *