FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2073544 · Received May 2, 2011

Report

Report Number
6000001-2011-03339
Event Type
Malfunction
Date Received
May 2, 2011
Date of Event
November 18, 2010
Report Date
April 18, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT DID CONFIRM THE SEPARATED TUBING CONDITION. THIS IS A SINGLE USE DEVICE AND WILL NOT BE REPAIRED. NO OTHER OBSERVATION WAS FOUND ON THE UNIT. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE WAS NOT IDENTIFIED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT OF ONE (1) CE INFUSOR, LV 2 ML/H FOR WHICH THE FLOW RESTRICTOR WAS SEPARATED FROM THE DELIVERY TUBING BEFORE OPENING THE OVERPOUCH. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 08D065

Patients

Seq Age Sex Outcome Treatment
1