FDA Adverse Event Injury Summary report: N

ALTRUA

MDR report key: 2073540 · Received May 2, 2011

Report

Report Number
2124215-2011-07588
Event Type
Injury
Date Received
May 2, 2011
Date of Event
January 17, 2011
Report Date
January 17, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE EXHIBITED LOSS OF CAPTURE AND FAILURE TO PACE. REPROGRAMMING CHANGES WERE DONE TO ALLEVIATE THESE ISSUES WHICH WERE UNSUCCESSFUL. AS THE PHYSICIAN WAS UNABLE TO DETERMINE IF THIS WAS A LEAD OR DEVICE ISSUE, THE LEAD WAS SURGICALLY ABANDONED AND THE DEVICE EXPLANTED. BOTH WERE SUCCESSFULLY REPLACED WITHOUT INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALTRUA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND S203

Patients

Seq Age Sex Outcome Treatment
1 89 YR Required Intervention 1688T| 1688TC| S203