EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0
Report
- Report Number
- 1038671-2024-04445
- Event Type
- Injury
- Date Received
- November 20, 2024
- Date of Event
- July 25, 2023
- Report Date
- December 12, 2024
- Manufacturer
- EXACTECH, INC.
- Product Code
- KWT
- UDI-DI
- 10885862086419
- PMA / PMN Number
- K063569
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
H3: THE REVISION REPORTED WAS LIKELY THE RESULT OF INCOMPLETE SEATING OF THE LINER DURING IMPLANTATION, BONE IMPINGEMENT, PATIENT-RELATED CONDITIONS, AN UNREPORTED POST-TRAUMATIC EVENT, OR ANY COMBINATION OF THESE POSSIBILITIES, WHICH LED TO HUMERAL LINER DISASSOCIATION. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT RETURNED FOR EVALUATION AND RADIOGRAPHS WERE NOT PROVIDED.
H3) PENDING EVALUATION (D10) CONCOMITANT DEVICE(S): GLENOSPHERE 42MM 320-01-42 6154427 HUMERAL LINER 42MM, +2.5MM 320-42-03 4483543 GLENOSPHERE LOCKING SCREW 320-15-05 6113795 REVERSE SHOULDER TORQUE DEFINING SCREW 320-20-00 6115379 RS GLENOID PLATE SUP AUG, 10 DEG 320-15-02 5958989 EQUINOXE PRESERVE STEM 11MM 300-30-11 5729693 EQ REV COMPRESS SCREW LCK CAP KIT, 4.5 X 38MM 320-20-38 5930239 EQ REV COMPRESS SCREW LCK CAP KIT, 4.5 X 38MM 320-20-38 6113880 EQ REV COMPRESS SCREW LCK CAP KIT, 4.5 X 26MM 320-20-26 6015034 EQ REV COMPRESS SCREW LCK CAP KIT, 4.5 X 42MM 320-20-42 6058870.
AS REPORTED, APPROXIMATELY 4 YEARS POST OP INITIAL RIGHT TSA, THIS 71 Y/O MALE PATIENT WAS REVISED DUE TO HUMERAL LINER DISSOCIATION. HUMERAL LINER WAS DISSOCIATED FROM THE HUMERAL TRAY. GLENOSPHERE, LINER AND TRAY WERE REVISED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. NO OTHER PATIENT INFORMATION/MEDICAL HISTORY REPORTED. PHOTO ATTACHED. UNABLE TO OBTAIN X-RAYS. PRODUCT NOT RETURNING - DISPOSED BY HOSPITAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2053862 | EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0 | PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED | KWT | EXACTECH, INC. | 10885862086419 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Male | Required Intervention| H | SEE H11. |