FDA Adverse Event Malfunction Summary report: N

ALINITY I FREE T4 REAGENT KIT

MDR report key: 20734841 · Received November 20, 2024

Report

Report Number
3005094123-2024-00602
Event Type
Malfunction
Date Received
November 20, 2024
Date of Event
September 2, 2024
Report Date
January 28, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
CEC
UDI-DI
00380740121822
PMA / PMN Number
K173122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. AN INCREASE IN COMPLAINTS HAS BEEN OBSERVED FOR LOT 60071UD00, HOWEVER, IN-HOUSE PERFORMANCE TESTING WAS COMPLETED WHICH INDICATES THE PRODUCT IS PERFORMING AS EXPECTED. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES, OR DEVIATIONS ASSOCIATED WITH LOT 60071UD00 AND THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ALINITY I FREE T4 REAGENT FOR LOT 60071UD00 WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

COMPLETE ENTRY FOR SECTION A1 - PATIENT IDENTIFIER: (B)(6). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR MULTIPLE PATIENTS. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE: FREE T4, 0.7 TO 1.48 NG/DL): ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT >5.00 NG/DL; REPEAT FT4 RESULT= 1.07 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT 3.5 NG/DL; REPEAT FT4 RESULT= 1.02 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT >5.0 NG/DL; REPEAT FT4 RESULT= 1.1 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 2.14 NG/DL; REPEAT FT4 RESULT= 1.08 NG/DL, ON (B)(6)2024. SID (B)(6), INITIAL FT4 RESULT= 1.67 NG/DL; REPEAT FT4 RESULT= 1.31 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.55 NG/DL; REPEAT FT4 RESULT= 1.26 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT >5.00 NG/DL; REPEAT FT4 RESULT= 1.24 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.64 NG/DL; REPEAT FT4 RESULT= 0.74 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.54 NG/DL; REPEAT FT4 RESULT= 0.9 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.54 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.56 NG/DL; REPEAT FT4 RESULT= 1.08 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 2.23 NG/DL; REPEAT FT4 RESULT= 1.48 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT >5.00 NG/DL; REPEAT FT4 RESULT= 0.89 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.76 NG/DL; REPEAT FT4 RESULT= 0.94 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.61 NG/DL; REPEAT FT4 RESULT= 1.32 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 3.07 NG/DL; REPEAT FT4 RESULT= 1.03 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.56 NG/DL; REPEAT FT4 RESULT= 1.16 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.84 NG/DL; REPEAT FT4 RESULT= 0.6 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.71 NG/DL; REPEAT FT4 RESULT= 1.05 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT >5.00 NG/DL; REPEAT FT4 RESULT= 1.14 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.6 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.62 NG/DL; REPEAT FT4 RESULT= 1.27 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 2.44 NG/DL; REPEAT FT4 RESULT= 0.92 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT >5.00 NG/DL; REPEAT FT4 RESULT= 0.87 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.58 NG/DL; REPEAT FT4 RESULT= 0.9 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.53 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.56 NG/DL; REPEAT FT4 RESULT= 0.99 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT >5.00 NG/DL; REPEAT FT4 RESULT= 1.08 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.51 NG/DL; REPEAT FT4 RESULT= 1.13 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.63 NG/DL; REPEAT FT4 RESULT= 1.15 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT 1.81 NG/DL; REPEAT FT4 RESULT= 1.19 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.53 NG/DL; REPEAT FT4 RESULT= 1.21 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.65 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.59 NG/DL; REPEAT FT4 RESULT= 0.9 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.77 NG/DL; REPEAT FT4 RESULT= 1.47 NG/DLON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.49 NG/DL; REPEAT FT4 RESULT= 1.03 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.75 NG/DL; REPEAT FT4 RESULT= 1.06 NG/DL, ON (B)(6)2024, SID (B)(6), INITIAL FT4 RESULT= 1.6 NG/DL; REPEAT FT4 RESULT= 0.94 NG/DL, ON (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.95 NG/DL; REPEAT FT4 RESULT= 1.33 NG/DL . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR MULTIPLE PATIENTS. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE: FREE T4, 0.7 TO 1.48 NG/DL): (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.00 NG/DL; REPEAT FT4 RESULT= 1.07 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 3.5 NG/DL; REPEAT FT4 RESULT= 1.02 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.0 NG/DL; REPEAT FT4 RESULT= 1.1 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 2.14 NG/DL; REPEAT FT4 RESULT= 1.08 NG/DL. (B)(6) 2024. SID (B)(6), INITIAL FT4 RESULT= 1.67 NG/DL; REPEAT FT4 RESULT= 1.31 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.55 NG/DL; REPEAT FT4 RESULT= 1.26 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.00 NG/DL; REPEAT FT4 RESULT= 1.24 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.64 NG/DL; REPEAT FT4 RESULT= 0.74 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.54 NG/DL; REPEAT FT4 RESULT= 0.9 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.54 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.56 NG/DL; REPEAT FT4 RESULT= 1.08 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 2.23 NG/DL; REPEAT FT4 RESULT= 1.48 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.00 NG/DL; REPEAT FT4 RESULT= 0.89 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.76 NG/DL; REPEAT FT4 RESULT= 0.94 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.61 NG/DL; REPEAT FT4 RESULT= 1.32 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 3.07 NG/DL; REPEAT FT4 RESULT= 1.03 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.56 NG/DL; REPEAT FT4 RESULT= 1.16 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.84 NG/DL; REPEAT FT4 RESULT= 0.6 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.71 NG/DL; REPEAT FT4 RESULT= 1.05 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.00 NG/DL; REPEAT FT4 RESULT= 1.14 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.6 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.62 NG/DL; REPEAT FT4 RESULT= 1.27 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 2.44 NG/DL; REPEAT FT4 RESULT= 0.92 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.00 NG/DL; REPEAT FT4 RESULT= 0.87 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.58 NG/DL; REPEAT FT4 RESULT= 0.9 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.53 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.56 NG/DL; REPEAT FT4 RESULT= 0.99 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 5.00 NG/DL; REPEAT FT4 RESULT= 1.08 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.51 NG/DL; REPEAT FT4 RESULT= 1.13 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.63 NG/DL; REPEAT FT4 RESULT= 1.15 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.81 NG/DL: REPEAT FT4 RESULT= 1.19 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.53 NG/DL; REPEAT FT4 RESULT= 1.21 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.65 NG/DL; REPEAT FT4 RESULT= 1.34 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.59 NG/DL; REPEAT FT4 RESULT= 0.9 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.77 NG/DL; REPEAT FT4 RESULT= 1.47 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.49 NG/DL; REPEAT FT4 RESULT= 1.03 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.75 NG/DL; REPEAT FT4 RESULT= 1.06 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.6 NG/DL; REPEAT FT4 RESULT= 0.94 NG/DL. (B)(6) 2024, SID (B)(6), INITIAL FT4 RESULT= 1.95 NG/DL; REPEAT FT4 RESULT= 1.33 NG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2460250 ALINITY I FREE T4 REAGENT KIT RADIOIMMUNOASSAY, FREE THYROXINE CEC ABBOTT IRELAND DIAGNOSTICS DIVISION 60071UD00 00380740121822

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I FREE T4 1200T, 07P70-30, 57607UD00| ALNTY I FREE T4 1200T, 07P70-30, 57607UD00| ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01,(B)(6)