START-X TIP EMS INSERT 3
Report
- Report Number
- 8031010-2024-00144
- Event Type
- Malfunction
- Date Received
- November 20, 2024
- Report Date
- November 26, 2024
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- ELC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EZ
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT RECEIVED WITH A SIMILAR DEVICE WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE LOT NUMBER WAS PROVIDED AND RETAINED-PRODUCT TESTING AND/OR DHR REVIEW ARE PLANNED. THE RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED START-X TIP EMS INSERT 3 IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1833636). WE CANNOT RULE ON THE COMPLIANCE OF THE POWER SETTINGS SELECTED BY THE CUSTOMER IN COMPARISON WITH THE ONES RECOMMENDED BY MAILLEFER WHICH ARE EXCLUSIVELY GIVEN FOR EMS GENERATORS. ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. GENERALLY SPEAKING, SEVERAL FACTORS MAY CONTRIBUTE TO BREAKAGE ISSUE, SUCH AS USURY, PRESSURE OR INCORRECT TECHNIQUE. ALL OF THESE FACTORS MAY AFFECT THE LONGEVITY OF THE TIP. ROOT CAUSES ARE NOT IDENTIFIED.
IN THIS EVENT IT IS REPORTED THAT START-X TIP EMS INSERT 3 BROKE DURING USE. ALL BROKEN PARTS WERE RETRIEVED FROM PATIENT'S MOUTH. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2313440 | START-X TIP EMS INSERT 3 | SCALER, ULTRASONIC | ELC | MAILLEFER INSTRUMENTS HOLDING SARL | 1833636 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |