ARCHITECT TOTAL PSA
Report
- Report Number
- 1628664-2011-00211
- Event Type
- Injury
- Date Received
- May 1, 2011
- Date of Event
- April 5, 2011
- Report Date
- April 7, 2011
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- LTJ
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE INITIAL REPORT WAS SUBMITTED UNDER BRAND NAME ABBOTT ARCHITECT I1000 ANALYZER, IRVING TX MANUFACTURING SITE. IT WAS DETERMINED THAT THE INVESTIGATION SHOULD BE PERFORMED ON BRAND NAME ARCHITECT TOTAL PSA REAGENT, (B)(4). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7K70 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 6C06. AS THE LOT NUMBER OF THE ARCHITECT TOTAL PSA REAGENT WAS UNKNOWN AND NO RECORDS REMAIN ON THE INSTRUMENT A SIX MONTH REVIEW OF THE ARCHITECT TOTAL PSA QUALITY TESTING DATA WAS PERFORMED ((B)(6) 2010 - (B)(6) 2011). THE RECORDS WERE ANALYZED AND NO PERFORMANCE ISSUES WERE IDENTIFIED. IN ADDITION, THERE WERE NO TRENDS OBSERVED. CUSTOMER COMPLAINT DATA WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED. THE ARCHITECT TOTAL PSA REAGENT PACKAGE INSERT WAS REVIEWED AND WAS FOUND TO ADEQUATELY ADDRESS THE ISSUE OF ERRATIC RESULTS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT DEFICIENCY.
(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE ACCOUNT STATED THAT THE ARCHITECT I1000 HAS GENERATED SEVERAL ELEVATED TOTAL PSA RESULTS ON A PATIENT. ON (B)(6) 2010 A TOTAL PSA RESULT OF 115.4 NG/ML WAS GENERATED AND A FREE PSA RESULT WAS 3.9%. ON (B)(6) 2011 A TOTAL PSA RESULT OF 113.6 NG/ML WAS GENERATED. ON (B)(6) 2011 A TOTAL PSA RESULT OF 124.5 NG/ML WAS GENERATED. THE PATIENT UNDERWENT A BIOPSY AS A RESULT OF THE ELEVATED PSA RESULTS. THE BIOPSY RESULTS WERE NEGATIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT TOTAL PSA | FOR THE DETERMINATION OF TOTAL PROSTATE SPECIFIC ANTIGEN (PSA) | LTJ | ABBOTT LABORATORIES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Disability | ARCHITECT I1000SR 1L86-01 (B)(4)| ARCHITECT TOTAL PSA LN 7K70 LOT UNKNOWN |