FDA Adverse Event Malfunction Summary report: N

OTHER

MDR report key: 2073331 · Received May 1, 2011

Report

Report Number
2531779-2011-03001
Event Type
Malfunction
Date Received
May 1, 2011
Report Date
April 1, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 06/06/2011 - DEVICE EVALUATION: THE CARTRIDGE HAS NOT BEEN RETURNED. A RETAINED SAMPLE FROM THE SAME LOT WAS EVALUATED BY PRODUCT ANALYSIS ON 05/10/2011 WITH THE FOLLOWING FINDINGS: THE CARTRIDGE PASSES VISUAL INSPECTION; THERE WAS NO DAMAGE OR DEFECTS NOTED TO THE LUER CONNECTION OR O-RINGS. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIRTY-THREE CARTRIDGES FROM LOT # B201609 HAVE BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE CARTRIDGES WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. LEAK TESTING WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

BRAND NAME: ANIMAS INSULIN CARTRIDGE. EXPIRATION: 10/31/2012. THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS TO REPORT LEAKING CARTRIDGES ON CURRENT LOT NUMBER. THE PATIENT REPORTED LEAKING AT THE HUB WHERE NEEDLE ATTACHES TO CARTRIDGE. THE PATIENT CLAIMED THE NEEDLE IS SECURE AND HAS NEVER HAPPENED BEFORE. THE PATIENT DENIED LEAKAGE AT SITE OR WHEN TUBING IS ATTACHED TO CARTRIDGE. THE PATIENT DENIED VISIBLE CRACKS IN FILL NEEDLE OR CARTRIDGE. THE PATIENT CLAIMED HE HAS USED CARTRIDGES FROM OTHER LOT# AND HAS NOT HAD AN ISSUE. THERE WAS NO MENTION OF SYMPTOMS OR MEDICAL TREATMENT REQUIRED AS A RESULT OF THE ALLEGED COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OTHER INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART

Patients

Seq Age Sex Outcome Treatment
1