FDA Adverse Event Death Summary report: N

SPRINT FIDELIS

MDR report key: 2073316 · Received May 1, 2011

Report

Report Number
2649622-2011-07371
Event Type
Death
Date Received
May 1, 2011
Date of Event
February 13, 2007
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0070-2008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE DEVICE AND LEADS ARE IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION. (B)(4): DEVICE WAS RETURNED AND ANALYZED AND FOUND TO HAVE NORMAL BATTERY DEPLETION. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

AN IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) AND TWO LEADS WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURE'S DATA BASE INDICATION THE PATIENT DIED APPROXIMATELY NINE MONTHS POST THE IMPLANT OF THE ICD.

Description of Event or Problem · 1

AN IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) AND TWO LEADS WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURE'S DATA BASE INDICATION THE PATIENT DIED APPROXIMATELY NINE MONTHS POST THE IMPLANT OF THE ICD. CAUSE OF DEATH WILL BE REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949 ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Death