FDA Adverse Event Malfunction Summary report: N

MAQUET POWERLED II

MDR report key: 20732933 · Received November 20, 2024

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
9710055-2024-0000597
Event Type
Malfunction
Date Received
November 20, 2024
Date of Event
October 22, 2024
Report Date
November 19, 2024
Manufacturer
MAQUET SAS
Product Code
FTD
UDI-DI
03700712426279
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

EVENT SITE NAME: (B)(6). ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 0

THE CORRECTION OF H4 MANUFACTURE DATE, D4 CATALOG #, UNIQUE IDENTIFIER (UDI) #, SERIAL#, VERSION OR MODEL #, B5 DESCRIBE EVENT AND PROBLEM, D1 BRAND NAME, D2A AND D2B PRODUCT CODE., B5 DESCRIBE EVENT AND PROBLEM DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS H4 MANUFACTURE DATE: 2021-10-07 CORRECTED H4 MANUFACTURE DATE: 2023-10-19 PREVIOUS D1 BRAND NAME EQT CORRECTED D1 BRAND NAME MAQUET POWERLED II PREVIOUS D2A PRODUCT CODE. FXR CORRECTED D2A PRODUCT CODE. FTD PREVIOUS D2B PRODUCT CODE DESCRIPTION. HOLDER, CAMERA, SURGICAL CORRECTED D2B PRODUCT CODE DESCRIPTION. LAMP, SURGICAL PREVIOUS D4 CATALOG # ARD567711940 CORRECTED D4 CATALOG # BLANK PREVIOUS D4 UNIQUE IDENTIFIER (UDI) #: N/A CORRECTED D4 UNIQUE IDENTIFIER (UDI) #: (B)(4). PREVIOUS D4 SERIAL # (B)(6). CORRECTED D4 SERIAL # (B)(6). PREVIOUS D4 VERSION OR MODEL # ARD567711940 CORRECTED D4 VERSION OR MODEL # ARDPWT639073A PREVIOUS B5 DESCRIBE EVENT AND PROBLEM: GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - EQT. IT WAS STATED THE MONITOR BOX FELL OFF DURING SUGERY, WHEN NURSE MOVED THE MONITOR. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS MONITOR FALLING OFF DURING PROCEDURE MAY CAUSE CONTAMINATION OF SURGERY TOOLS OR CAN CAUSE SERIOUS INJURY TO MEDICAL STAFF. CORRECTED B5 DESCRIBE EVENT AND PROBLEM: GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS ¿ POWERLED II. IT WAS STATED THE MONITOR BOX FELL OFF DURING SURGERY, WHEN NURSE MOVED THE MONITOR. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS MONITOR FALLING OFF DURING PROCEDURE MAY CAUSE CONTAMINATION OF SURGERY TOOLS OR CAN CAUSE SERIOUS INJURY TO MEDICAL STAFF. GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS ¿ MAQUET POWERLED II. IT WAS STATED THE MONITOR BOX FELL OFF DURING SURGERY, WHEN NURSE MOVED THE MONITOR. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS MONITOR FALLING OFF DURING PROCEDURE MAY CAUSE CONTAMINATION OF SURGERY TOOLS OR CAN CAUSE SERIOUS INJURY TO MEDICAL STAFF. BASED ON THE INFORMATION COLLECTED, IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE SURGICAL LIGHT DID NOT MEET ITS SPECIFICATION AND IN THIS WAY THE DEVICE CONTRIBUTED TO EVENT. THE CLAIMED DEVICE WAS BEING USED FOR PATIENT TREATMENT WHEN THE EVENT TOOK PLACE. A ROOT CAUSE ANALYSIS WAS CONDUCTED BY THE SUBJECT MATTER EXPERT. AS STATED: THE REAR BOX 01 MAY NOT HAVE BEEN SECURED PROPERLY DURING THE INSTALLATION OR A MAINTENANCE. IT CAN BE CONSIDERED AS AN ISOLATED CASE. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DOES NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.

Description of Event or Problem · 0

GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL EQUIPMENT - EQT. IT WAS STATED THE MONITOR BOX FELL OFF DURING SUGERY, WHEN NURSE MOVED THE MONITOR. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS MONITOR FALLING OFF DURING PROCEDURE MAY CAUSE CONTAMINATION OF SURGERY TOOLS OR CAN CAUSE SERIOUS INJURY TO MEDICAL STAFF.

Description of Event or Problem · 0

GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS ¿ POWERLED II. IT WAS STATED THE MONITOR BOX FELL OFF DURING SURGERY, WHEN NURSE MOVED THE MONITOR. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS MONITOR FALLING OFF DURING PROCEDURE MAY CAUSE CONTAMINATION OF SURGERY TOOLS OR CAN CAUSE SERIOUS INJURY TO MEDICAL STAFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1995697 MAQUET POWERLED II LAMP, SURGICAL FTD MAQUET SAS ARDPWT639073A 03700712426279

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown