FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2073177 · Received April 29, 2011

Report

Report Number
6000001-2011-03304
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
April 11, 2011
Report Date
April 11, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT CONFIRMED THE REPORTED CONDITION OF A LEAK, OBSERVED AT THE CONNECTION OF THE BLUE WINGED LUER CAP. THE ROOT CAUSE WAS DETERMINED TO BE USER ERROR; THE CAP WAS NOT SECURELY TIGHTENED. AFTER THE CAP WAS SECURELY TIGHTENED AND THE CLAMP CLOSED, THE LEAKAGE COMPLETELY STOPPED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE AND/OR ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

BAXTER (B)(4) PRODUCT SURVEILLANCE RECEIVED A COMPLAINT THAT ONE (1) CE INTERMATE LV 100 WAS LEAKING AT THE BLUE WINGED CAP AFTER BEING FILLED WITH PAMIDRONATE. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 10N008

Patients

Seq Age Sex Outcome Treatment
1 PAMIDRONATE