ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2011-01804
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- November 26, 2010
- Report Date
- April 28, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION NOT PROVIDED ON INITIAL CONTACT.
(B)(4). THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. THE INSTRUMENT DID NOT HAVE A BROKEN BLADE WHEN RECEIVED. IT IS POSSIBLE THAT THE DEVICE RETURNED MAY HAVE BEEN USED TO COMPLETE THE PROCEDURE AND IS NOT THE DEVICE COMPLAINED ABOUT. DURING FUNCTIONAL TESTING, AN ERROR CODE 5 WAS DISPLAYED AND THE BLADE TIP FURTHER CRACKED. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. WHEN A BLADE HAS BEEN COMPROMISED SUCH AS SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. A PROBABLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC RECTOSIGMOIDECTOMY TO DEEP ENDOMETRIOSIS PROCEDURE, THE SHEARS PRESENTED THE BREAK OF THE ACTIVE SPADE BEING WITH THAT NECESSARY THE EXCHANGE OF THE SHEARS FOR THE CONCLUSION OF THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |