FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 2073135 · Received April 29, 2011

Report

Report Number
6000001-2011-03299
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
April 11, 2011
Report Date
April 11, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THE DEVICE HAS NOT BEEN PREVIOUSLY SENT INTO SERVICE FOR THE REPORTED CONDITION OF "FAILURE CODE 550:320:844:0000 FAILED TO AUDIBLY ALARM."

Additional Manufacturer Narrative · 1

THE CONDITION OF FAILED TO AUDIBLY ALARM WAS CONFIRMED THROUGH EVALUATION AND WAS FOUND TO BE DUE TO A DISCONNECTED REAR HARNESS CONNECTOR. THE REAR HARNESS CONNECTOR WAS RECONNECTED TO CORRECT THE CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

DURING DEVICE EVALUATION BY BAXTER ON A COLLEAGUE INFUSION PUMP FAILURE CODE 550:320:844:0000 FAILED TO AUDIBLY ALARM. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS DEVICE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.04.00 CATEGORIZED AS UNREMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1