FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2073124 · Received March 23, 2011

Report

Report Number
2531779-2011-01885
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
February 25, 2011
Report Date
February 25, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO OBTAIN A RESPONSE DURING TESTING. IT WAS OBSERVED THAT ALL BUTTON KEY CONTACTS DO NOT SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE UNRESPONSIVE. THE PT REPORTED THAT THE DOWN ARROW KEYPAD BUTTON IS UNRESPONSIVE. SHE NOTED THAT THE KEYPAD BUTTONS DO NOT SPRING BACK WHEN PRESSED. THE PT STATED THAT THE RUBBER KEYPAD IS INTACT, THE PUMP IS EXPOSED TO WATER WHEN SHE SWIMS, AND SHE WEARS THE PUMP ON HER WAIST WITH A CLIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 32 YR