FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2073120 · Received March 23, 2011

Report

Report Number
2531779-2011-01838
Event Type
Malfunction
Date Received
March 23, 2011
Report Date
February 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOW NON-ZERO FORCE WAS DETECTED ASSOCIATED WITH LOSS OF PRIME ALARMS, WHICH WERE DUPLICATED DURING INVESTIGATION.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1