FDA Adverse Event Malfunction Summary report: N

LIFEPAK 15 MONITOR/DEFIBRILLATOR

MDR report key: 2073114 · Received March 17, 2011

Report

Report Number
3015876-2011-00246
Event Type
Malfunction
Date Received
March 17, 2011
Date of Event
February 18, 2011
Report Date
February 18, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K082937
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE SYSTEM PCB ASSEMBLY AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PATIENT CARE, THE DEVICE WOULD NOT POWER ON. THE END USER ATTEMPTED AGAIN WITH A DIFFERENT SET OF BATTERIES WITH THE SAME RESULT. THERE WAS AN APPROXIMATE DELAY IN CARE OF 4-5 MINUTES WHEN THE PATIENT WAS ADMITTED TO THE LOCAL HOSPITAL. THERE WAS NO ADVERSE OUTCOME TO THE PATIENT FROM RESULT OF THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 15 MONITOR/DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. 15 NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR