FDA Adverse Event Malfunction Summary report: N

COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 2073095 · Received April 29, 2011

Report

Report Number
6000001-2011-03294
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
January 1, 2011
Report Date
January 14, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). UPON FURTHER REVIEW OF THE EVENT HISTORY, THE OCCURRENCE DATE OF THIS EVENT WAS DETERMINED TO BE (B)(6) 2010. EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:03 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY BY A BAXTER SERVICE TECHNICIAN. THIS CONDITION WAS CAUSED BY A FAULTY CHANNEL B AIR IN LINE (AIL) PRINTED CIRCUIT BOARD (PCB). THE CHANNEL B PUMPHEAD MODULE WAS REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:03 ON CHANNEL B, WHICH OCCURRED IN THE GENERAL PT WARD. UPON REVIEWING THE PUMP'S EVENT HISTORY, BAXTER QUALITY ENGINEERING DISCOVERED THIS EVENT INTERRUPTED DELIVERY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1