FDA Adverse Event Injury Summary report: N

TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET, MODELS 1086, 1087, 1088

MDR report key: 2072997 · Received April 29, 2011

Report

Report Number
3005099803-2011-01420
Event Type
Injury
Date Received
April 29, 2011
Date of Event
April 8, 2011
Report Date
April 8, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
LQC
PMA / PMN Number
K040447
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT INDICATED THAT THE DEVICE WAS DISPOSED AND WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET WAS USED DURING AN ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY) PROCEDURE PERFORMED ON (B)(6), 2011. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SOLID WIRE WITHIN THE HANDLE BROKE WHILE USING THE ALLIANCE HANDLE TO CRUSH A STONE. AFTER THE WIRE BROKE THE HANDLE WAS REMOVED AND THE WIRE WAS ATTACHED TO A SOEHENDRA LITHOTRIPSY SYSTEM AND THE PHYSICIAN WAS ABLE TO 'POP' THE TIP OF THE BASKET AND REMOVE THE DEVICE FROM THE PATIENT. THE PANCREATIC STONE WAS APPROXIMATELY 1.6CM AND IT REMAINS WITHIN THE PATIENT ALONG WITH THE DETACHED DEVICE TIP. THE PROCEDURE LASTED 2-3 HOURS AND THE PATIENT SPENT ONE NIGHT IN THE HOSPITAL. IT WAS REPORTED THAT THE PATIENT IS IN WELL CURRENT CONDITION AND WILL BE RESCHEDULED FOR A LASER LITHOTRIPSY AT A FUTURE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET, MODELS 1086, 1087, 1088 LITHOTRIPTOR, BILIARY MECHANICAL LQC BOSTON SCIENTIFIC - SPENCER M00510870 14181760

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| O