ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-02749
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- December 26, 2010
- Report Date
- March 26, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. IT WAS OBSERVED THAT THE KEYPAD BUTTON SYMBOLS ARE WORN. EVAL CONFIRMED THAT THE KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS.
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. A FAMILY MEMBER REPORTED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS HAVE BEEN INTERMITTENTLY UNRESPONSIVE FOR THE PAST 3 MONTHS. SHE NOTED THAT THE OK, UP ARROW, AND DOWN ARROW BUTTON SYMBOLS ARE WORN, AND THAT THE BUTTONS STILL CLICK AND SPRING BACK WHEN PRESSED. THE FAMILY MEMBER CONFIRMED THAT THE KEYPAD IS INTACT; DENIED EXPOSING THE PUMP TO WATER; AND STATED THAT THE PT WEARS THE PUMP ON HER WAIST WITH A CLIP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |