FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2072946 · Received April 3, 2011

Report

Report Number
2531779-2011-02139
Event Type
Malfunction
Date Received
April 3, 2011
Report Date
March 3, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. DURING INVESTIGATION, IT WAS FOUND THAT ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ARE STICKING AND VERY HARD TO PRESS TO PROGRAM CORRECTLY. IT WAS ALSO REPORTED THAT THERE IS NO STRUCTURAL DAMAGE TO KEYPAD AND NO WATER EXPOSURE TO PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1