FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2072944 · Received April 2, 2011

Report

Report Number
2531779-2011-02136
Event Type
Malfunction
Date Received
April 2, 2011
Report Date
March 3, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. IT WAS ALSO FOUND THAT THE KEYPAD SYMBOLS ARE WORN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ARE GETTING SOFT. IT WAS REPORTED THAT THE KEYPAD IS INTACT AND THE PUMP IS NOT EXPOSED TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment