FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2072942 · Received April 2, 2011

Report

Report Number
2531779-2011-02127
Event Type
Malfunction
Date Received
April 2, 2011
Report Date
March 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. ALL KEYPAD BUTTON PRESSES REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE DURING TESTING. DURING INVESTIGATION, IT WAS FOUND THAT ALL KEYPAD BUTTONS INTERMITTENTLY SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD IS INTERMITTENTLY RESPONDING. IT WAS REPORTED THERE ARE NO TEARS OR PEELING OF THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1