FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 2072911 · Received April 29, 2011

Report

Report Number
3005075853-2011-01789
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
February 8, 2011
Report Date
March 1, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). LOCKOUT. ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RETURNED EMPTY AND WITH THE LOCK OUT MECHANISM BROKEN AS IT WAS FIRED THROUGH. THE INSTRUMENT IS DESIGNED TO LOCK OUT AFTER ALL THE CLIPS HAVE BEEN FIRED; THEREFORE A POTENTIAL CAUSE OF THE CUSTOMER REPORTED EXPERIENCE IS THE FIRING OF ALL OF THE CLIPS AND THE INSTRUMENT "WOULD NOT FIRE" (ACTIVATION OF THE LOCK OUT MECHANISM). THE INSTRUMENT HAS AN ORANGE INDICATOR THAT APPEARS ON THE TOP OF THE HANDLE AS A REFERENCE FOR THE USER AS TO THE QUANTITY OF CLIPS REMAINING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, TWO DEVICES JAMMED AND WOULD NOT FIRE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4U97D

Patients

Seq Age Sex Outcome Treatment
1