FDA Adverse Event Malfunction Summary report: N

CCU BED

MDR report key: 2072892 · Received March 31, 2011

Report

Report Number
1824206-2011-01966
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
March 9, 2011
Report Date
March 9, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE LOCKOUT BOARD TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED FUNCTIONS ARE NOT WORKING FROM THE SIDERAILS DUE TO FLUID INGRESS ONTO THE LOCKOUT BOARD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CCU BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 894

Patients

Seq Age Sex Outcome Treatment
1