FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2072891 · Received March 31, 2011

Report

Report Number
1824206-2011-01963
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
March 12, 2011
Report Date
March 12, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND BED IN MEDSURG ROOM. PATIENT WAS ABLE TO PLACE NURSE CALL, RESPONSE NOT HEARD IN EXPECTED LOCATION. CAUSE: (B)(4) CABLE WAS NOT CONNECTED. RECONNECTED (B)(4) CABLE TESTED FUNCTIONS TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINT ALLEGED THAT PATIENT WAS NOT ABLE TO PLACE NURSE CALL. RESPONSE NOT HEARD IN EXPECTED LOCATION. NO INJURY ALLEGED BY ACCOUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. P1900

Patients

Seq Age Sex Outcome Treatment
1