FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 2072858 · Received April 29, 2011

Report

Report Number
3005075853-2011-01785
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
April 4, 2011
Report Date
April 6, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). LOCKOUT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. HOWEVER, THE LOCKOUT MECHANISM WAS FOUND TO BE NON-FUNCTIONAL. IN ORDER TO EVALUATE THE CONDITION OF THE DEVICE'S INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE HANDLE LOCKOUT POSTS WERE FOUND TO BE BROKEN WHICH DENOTES THAT THE LOCKOUT HAD BEEN FIRED THROUGH AS A RESULTS OF A LOCKOUT PREMATURE. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE FOUND CONDITION. PLEASE NOTE THAT THIS CONDITION IS UNRELATED WITH THE INCIDENT REPORTED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS WERE MALFORMED AND SLIPPED OUT OF THE JAW. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA F4PT0R

Patients

Seq Age Sex Outcome Treatment
1