FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2072852 · Received April 29, 2011

Report

Report Number
2024168-2011-02997
Event Type
Injury
Date Received
April 29, 2011
Date of Event
April 6, 2011
Report Date
April 6, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. A CUFF MISS CAN OCCUR DUE TO A NUMBER OF FACTORS INCLUDING, BUT NOT LIMITED TO, NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE, AGGRESSIVE AND FAST DEPLOYMENT OF THE PLUNGER OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT DURING NEEDLE DEPLOYMENT. THIS MAY HAVE BEEN A CONTRIBUTING FACTOR TO THIS EVENT, BUT CANNOT BE CONFIRMED. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY MANUFACTURING RELATED NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. BASED ON THE INFORMATION RECEIVED WITH THIS COMPLAINT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE CANNOT BE DETERMINED. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, A 100% IN PROCESS TESTING OF DEVICES ARE PERFORMED TO ASSURE THERE IS NO NEEDLE DEFLECTION, WHICH MAY CAUSE THE EVENT REPORTED IN THIS CASE. A SAMPLING OF FINISHED DEVICES IS ALSO TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, ADDITIONAL INFORMATION WAS NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 020136H

Patients

Seq Age Sex Outcome Treatment
1 46 YR Required Intervention