FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXI 600 ACCESS® IMMUNOASSAY SYSTEM WITH DUAL GANTRY
MDR report key: 2072824
·
Received April 29, 2011
Report
- Report Number
- 2122870-2011-01190
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- March 29, 2011
- Report Date
- March 31, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FIELD SERVICE ENGINEER (FSE) REPLACED THE BOARD TO RESOLVE THE ISSUE. FSE VERIFIED THE REPAIR PER ESTABLISHED PROCEDURES. ROOT CAUSE IS ASSOCIATED WITH THE HARDWARE.
Description of Event or Problem · 1
WHEN A BECKMAN COULTER INC. (BCI) FIELD SERVICE ENGINEER (FSE) WAS PERFORMING AN INSTALL OF THE UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM WITH DUAL GANTRY , THE REAGENT STORAGE WAS UNABLE TO INITIALIZE. FSE FOUND A BURNED COMPONENT ON THE REAGENT STORAGE PCB. THERE WAS NO REPORT OF A BURNING OR ELECTRICAL SMELL AT THE TIME OF EVENT. THERE WERE NO INJURY, NO SMOKE, OR FLAMES. THE FIRE DEPARTMENT WAS NOT DISPATCHED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXI 600 ACCESS® IMMUNOASSAY SYSTEM WITH DUAL GANTRY | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXI 600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |