FDA Adverse Event Malfunction Summary report: N

UNICEL® DXI 600 ACCESS® IMMUNOASSAY SYSTEM WITH DUAL GANTRY

MDR report key: 2072824 · Received April 29, 2011

Report

Report Number
2122870-2011-01190
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
March 29, 2011
Report Date
March 31, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K023764
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FIELD SERVICE ENGINEER (FSE) REPLACED THE BOARD TO RESOLVE THE ISSUE. FSE VERIFIED THE REPAIR PER ESTABLISHED PROCEDURES. ROOT CAUSE IS ASSOCIATED WITH THE HARDWARE.

Description of Event or Problem · 1

WHEN A BECKMAN COULTER INC. (BCI) FIELD SERVICE ENGINEER (FSE) WAS PERFORMING AN INSTALL OF THE UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM WITH DUAL GANTRY , THE REAGENT STORAGE WAS UNABLE TO INITIALIZE. FSE FOUND A BURNED COMPONENT ON THE REAGENT STORAGE PCB. THERE WAS NO REPORT OF A BURNING OR ELECTRICAL SMELL AT THE TIME OF EVENT. THERE WERE NO INJURY, NO SMOKE, OR FLAMES. THE FIRE DEPARTMENT WAS NOT DISPATCHED.