FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2072799 · Received March 31, 2011

Report

Report Number
3004209178-2011-80898
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
March 22, 2011
Report Date
March 22, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD BEEN EXPERIENCING ELEVATED BLOOD GLUCOSE LEVELS. THE CUSTOMER STATED THAT HIS LAST INFUSION SET CHANGE WAS THE MORNING OF THE EVENT AND HE USED A MODEL MMT-396 QUICK-SET INFUSION SET. PROGRAMMING WAS CORRECT AND TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP PASSED THE FIXED PRIME BUT FAILED THE HIGH PRESSURE TEST TWICE AND WAS STUCK IN A REWIND LOOP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 33 YR MODEL MMT-396, LOT 9200573.| UNOMEDICAL QUICK-SET INSULIN INFUSION SET: