FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2072797 · Received March 31, 2011

Report

Report Number
3004209178-2011-80894
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
March 21, 2011
Report Date
March 30, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP MOTOR WAS SEATED IN THE BEGINNING OF THE DISPLACEMENT TEST DUE TO A LOOSE MOTOR SUPPORT DISK. AS A RESULT, WE WERE UNABLE TO PERFORM THE DISPLACEMENT TEST.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A NO DELIVERY ALARM ON THE INSULIN PUMP. THE CUSTOMER'S MOTHER STATED THAT SHE IS HAVING A HARD TIME PRIMING THE INSULIN PUMP AFTER THE ALARM WAS CLEARED. TROUBLESHOOTING WAS PERFORMED. NUMBERS STARTED RAMPING UP ON THE INSULIN PUMP WHILE IT WAS COMPLETELY EMPTY AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 19 YR