LIGAMAX-5MM
Report
- Report Number
- 3005075853-2011-01781
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- February 4, 2011
- Report Date
- February 4, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADVANCER. ADDITIONAL INFORMATION WAS REQUESTED. THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH JAWS IN THE CLOSED POSITION. THE TRIGGER WAS MANUALLY ASSISTED IN ORDER TO OPEN THE JAWS WITHOUT ANY DIFFICULTIES NOTED; ONE PEAR SHAPED CLIP WAS RELEASED. UPON FIRING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED; NO SCISSORED CLIPS WERE NOTED. A POSSIBLE CAUSE FOR THE MALFORMED CLIP MAY BE THAT AN ADVANCER BYPASS HAPPENED. THIS CAN OCCUR WHEN THE TIP OF THE ADVANCER IS ABOVE OR BELOW THE CLIP INSTEAD OF BEING POSITIONED AT THE BACK OF THE CLIP, SINCE THE CLIP IS NOT FULLY ADVANCED IN THE JAWS. THIS WILL PREVENT THE JAWS FROM CLOSING PROPERLY, RESULTING IN A PEAR SHAPED CLIP AND SUBSEQUENTLY THE JAWS MAY REMAIN CLOSED AFTER FIRING, REQUIRING A MANUAL TRIGGER OPENING. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT VISIBLE. THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, ON THE FIRST FIRING THE CLIP SCISSORED AND THE DEVICE LOCKED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |