FDA Adverse Event Malfunction Summary report: N

STERLING ES PTA BALLOON DILATATION CATHETER

MDR report key: 2072756 · Received April 29, 2011

Report

Report Number
2134265-2011-01409
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
April 2, 2011
Report Date
April 4, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: OVER 18 YEARS OLD. DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MFR REPORT #: 2134265-2011-01414 IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE TWO BALLOON RUPTURES OCCURRED. THE 100% STENOSED, DE NOVO TARGET LESION WAS LOCATED IN THE CALCIFIED AND MODERATELY TORTUOUS LEFT ANTERIOR TIBIAL ARTERY. PRE-DILATION WAS ATTEMPTED WITH A 1.5X20MM STERLING BALLOON CATHETER WHICH WAS ADVANCED TO TREAT THE TARGET LESION. UPON THE FIRST INFLATION AT 4 ATM'S THE BALLOON RUPTURED. A SECOND 1.5X20MM STERLING BALLOON CATHETER WAS THEN ADVANCED AND UPON THE FIRST INFLATION AT 6 ATM'S, IT ALSO RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING ES PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939134152010 14134909

Patients

Seq Age Sex Outcome Treatment
1 INFLATION DEVICE: GOODTECH| INTRODUCER: 5FR MEDKIT 55 CM| GUIDE WIRE: 0.014INCH ASTAT 175 CM