FDA Adverse Event Malfunction Summary report: N

CONNECTA PLUS3 WHITE PEGS

MDR report key: 20727514 · Received November 19, 2024

Report

Report Number
9610847-2024-00353
Event Type
Malfunction
Date Received
November 19, 2024
Date of Event
October 30, 2024
Report Date
December 24, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMG
UDI-DI
00382903946013
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 394601 AND LOT NUMBER 4061448. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

SOME ITEMS SEEM TO BE BADLY WELDED/CONNECTED, AND FLUID LEAKED UNDER PRESSURE I HAVE A CUSTOMER WHO SENT ME BACK 1 TAP BOX REFERENCE 394601 LOT 4061448. IT WOULD BE BADLY WELDED AND WOULD CAUSE LEAKAGE WITH THE PRESSURE. NO PROBLEM ENCOUNTERED WITH ANOTHER BOX FROM THE SAME BATCH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1938505 CONNECTA PLUS3 WHITE PEGS STOPCOCK, I.V. SET FMG BECTON DICKINSON 4061448 00382903946013

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown