MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2011-00128
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- March 31, 2011
- Report Date
- March 31, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K002489
- Removal / Correction Number
- 2955842-051613-005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS MDR IS BEING SUBMITTED FOR RETROSPECTIVE ACTIVITY PERFORMED RELATING TO FIELD ACTION NUMBER 2955842-051613-005 TO INVESTIGATE MICRO-CRACKS ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT; HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESES MICRO-CRACKS IS CONFINED TO 2 CM OF THE DISTAL END OF THE INSTRUMENT SHAFT. THIS INSTRUMENT WAS INSPECTED AS PART OF THIS RETROSPECTIVE ACTIVITY AND FOUND TO CONTAIN CRACKS ON THE INSTRUMENT MAIN TUBE.
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND A SEGMENT OF THE RIGHT BLADE TO BE BROKEN OFF FROM THE INSTRUMENT. THE BROKEN PIECE WAS NOT RETURNED. THE BLADE FRACTURED AT THE CROSS SECTION OF THE GRIP CABLE CRIMP AND HALF OF THE CABLE CRIMP IS EXPOSED. THE FRACTURED PLANE OF THE BLADE IS NEARLY STRAIGHT. NO OTHER DAMAGE WAS FOUND.
IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, THE 'JAW' OF THE MONOPOLAR CURVED SCISSORS INSTRUMENT PARTIALLY BROKE AND FELL INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH A REPLACEMENT INSTRUMENT OF THE SAME TYPE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.