HOMECHOICE
Report
- Report Number
- 1423500-2011-05189
- Event Type
- Malfunction
- Date Received
- April 29, 2011
- Date of Event
- April 7, 2011
- Report Date
- April 7, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE DEVICE FAILED THE RITE ELECTRICAL GROUND BOND TEST. THE ASSIGNABLE CAUSE FOR GROUND BOND WAS DETERMINED TO BE A LOOSE CONNECTION AT THE DOOR FRAME TO DOOR POST INTERFACE. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE GROUND BOND FAILURE. RENAL QUALITY ENGINEERING, ALONG WITH PLANT SERVICING AND QUALITY PERSONNEL, WILL CONTINUE TO MONITOR PRODUCT LINES FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THE GROUND BOND TEST INDICATING A HIGH RESISTANCE VALUE BETWEEN THE HC AND THE GROUND BOND ON THE POWER SUPPLY.