FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2072706 · Received March 31, 2011

Report

Report Number
2531779-2011-02115
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
February 16, 2011
Report Date
March 2, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. THE DOWN ARROW AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE DURING TESTING. THE UP ARROW AND CONTRAST KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO OBTAIN A RESPONSE. DURING INVESTIGATION, THE DOWN ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED. IT WAS OBSERVED THAT THE UP ARROW KEY CONTACT DOES NOT SPRING BACK WHEN PRESSED. THE DOWN ARROW, OK, AND CONTRAST KEY CONTACTS INTERMITTENTLY SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW KEYPAD BUTTON HAS REQUIRED MULTIPLE PRESSES TO OBTAIN A RESPONSE FOR THE PAST 2 WEEKS, AND TODAY THE UP ARROW KEYPAD BUTTON IS COMPLETELY UNRESPONSIVE. SHE NOTED THAT THE UP ARROW BUTTON DOES NOT CLICK WHEN PRESSED, BUT ALL OTHER KEYPAD BUTTONS FEEL NORMAL AND CLICK AS EXPECTED. THE PATIENT STATED THAT THE RUBBER KEYPAD IS INTACT; AND DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 57 YR