FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2072704 · Received March 31, 2011

Report

Report Number
2531779-2011-02130
Event Type
Malfunction
Date Received
March 31, 2011
Date of Event
February 1, 2011
Report Date
March 4, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL BUTTON KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS HAVE BEEN SLOW TO RESPOND FOR THE PAST FEW WEEKS. SHE NOTED THAT THE BUTTONS DO NOT CLICK AS EXPECTED WHEN PRESSED. THE PATIENT CONFIRMED THAT THE RUBBER KEYPAD IS INTACT; DENIED EXPOSING THE PUMP TO MOISTURE; AND STATED THAT SHE WEARS THE PUMP IN HER POCKET OR CLIPPED TO HER WAIST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 38 YR