FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2072641 · Received April 29, 2011

Report

Report Number
1423500-2011-05179
Event Type
Malfunction
Date Received
April 29, 2011
Date of Event
January 28, 2011
Report Date
January 28, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT FOR A CONNECTION ISSUE WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. DURING INVESTIGATION BY BAXTER THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. ROOT CAUSE WAS DETERMINED TO BE USE ERROR- INCOMPLETE CONNECTION. LABELING REVIEW FOUND LABELING TO BE ADEQUATE FOR THE USE ERROR IDENTIFIED IN THIS COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER, THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER AND REPORTED THAT THE CATHETER WAS NOT ALL THE WAY CONNECTED TO THE PATIENT LINE. THE PATIENT NOTICED THIS CONNECTION ISSUE DURING FILL 1. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HOME PATIENT (HP) TO CYCLE POWER OFF/ON AND START OVER WITH ALL NEW SUPPLIES AND MAKE SURE TO DRAIN FIRST. PRODUCT SURVEILLANCE CONTACTED THE PATIENT ON 04/28/2011. ACCORDING TO THE PATIENT, THERE WAS NO DEFECT WITH THE SUPPLIES. THE PATIENT STATED THAT IT WAS USE ERROR AS THE PATIENT DID NOT FULLY INSERT THE LINE. THE PATIENT STATED THAT EVERYTHING HAS BEEN FINE SINCE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 52 YR HOMECHOICE