FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 2072599 · Received April 25, 2011

Report

Report Number
9615739-2011-00101
Event Type
Malfunction
Date Received
April 25, 2011
Date of Event
March 29, 2011
Report Date
March 29, 2011
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE CASTER SUPPORTS WERE CRACKED AND THREE BRAKE CASTERS WOBBLED. HE REPLACED THE CASTERS AND CASTER SUPPORTS TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BRAKE WILL NOT HOLD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment