FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR PLUS DEFIBRILLATOR

MDR report key: 2072569 · Received April 21, 2011

Report

Report Number
3015876-2011-00341
Event Type
Malfunction
Date Received
April 21, 2011
Date of Event
March 25, 2011
Report Date
March 25, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K011144
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED NO AUDIO FAILURE. PHYSIO DETERMINED THE CAUSE OF THE MALFUNCTION TO BE DUE TO AN INTERMITTENT SPEAKER OUTPUT. A REPLACEMENT DEVICE WAS PROVIDED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD NO VOICE PROMPTS. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK CR PLUS DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. CRPLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA