ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2011-07566
- Event Type
- Injury
- Date Received
- April 29, 2011
- Date of Event
- March 21, 2011
- Report Date
- March 21, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A DEVICE CHECK IT WAS NOTED THAT THE PATIENT HAD 2300 VENTRICULAR FIBRILLATION EVENTS THAT WERE DIVERTED DUE TO NOISE ON THE RIGHT VENTRICULAR (RV) CHANNEL. PACING INHIBITION OF TWO SECONDS WAS ALSO OBSERVED DURING THE RECONFIRMATION PERIOD. A CHEST X-RAY REVEALED THAT ALL THREE LEADS (RV, RIGHT ATRIAL AND LEFT VENTRICULAR) WERE KINKED AS THEY CAME OUT OF THE SUBCLAVIAN VEIN. IT WAS NOTED THAT THE PATIENT WAS ALSO IN ATRIAL FLUTTER, HOWEVER THE BOSTON SCIENTIFIC SALES REPRESENTATIVE STATED THAT HE WAS UNCERTAIN IF THE FLUTTER WAS REAL OR NOISE. THE PACEMAKER DEPENDENT PATIENT HAD PRESENTED TO THE DEVICE CHECK WITH SHORTNESS OF BREATH AND LIGHTHEADEDNESS, BUT HAD NO FURTHER SYMPTOMS AND WAS UNAWARE OF ANY ISSUES GOING ON WITH THE SYSTEM. A LEAD REVISION PROCEDURE WAS PERFORMED AND THE RV LEAD WAS FOUND TO BE FRACTURED DUE TO THE PLACEMENT OF A SUTURE DURING THE ORIGINAL IMPLANT PROCEDURE. IT WAS NOTED THAT THE SUTURE WAS NOT PLACED ON THE SUTURE SLEEVE AND OVER TIME HAD RUBBED THROUGH THE LEAD INSULATION AND CONDUCTOR WIRES, FRACTURING THE LEAD. THE CHRONIC RV LEAD WAS SURGICALLY ABANDONED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED AS A RESULT OF THE LEAD REVISION PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0158 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Hospitalization| L| R | 4470| 0158| 0180| H135| N118| H215| 4518 |