PAXGENE® BLOOD DNA TUBE
Report
- Report Number
- 9617032-2024-01806
- Event Type
- Malfunction
- Date Received
- November 19, 2024
- Date of Event
- October 23, 2024
- Report Date
- December 5, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- PJE
- UDI-DI
- 04053727611659
- PMA / PMN Number
- K231469
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: MATERIAL #: 761165. LOT/BATCH #: 4095154. BD RECEIVED 4 PHOTOS FOR INVESTIGATION. THE PHOTO WERE REVIEWED AND THE INDICATED FAILURE MODE OF INCORRECT LABEL INFORMATION WAS OBSERVED. EVALUATION INDICATED THAT THE SHELF LABEL HAD A LOT NUMBER 4095154 WITH EXPIRY DATE 30TH OF APRIL 2025, AND TUBE LABELS OF LOT 4079048 WITH EXPIRY DATE 31ST OF MAY 2025. ADDITIONALLY, 200 RETENTION SAMPLES FROM BD INVENTORY, WERE EVALUATED BY VISUAL EXAMINATION AND UPON COMPLETION THE INDICATED FAILURE MODE FOR INCORRECT LABEL INFORMATION WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, SOME PRODUCT REQUIREMENTS FOR LOT RELEASE WERE UNMET AND NOT DISCOVERED PRIOR TO RELEASE. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE INCORRECT LABEL INFORMATION. A GLOBAL HOLD HAS BEEN PLACED ON THIS BATCH NUMBER. BD HAS INITIATED FURTHER ROOT CAUSE INVESTIGATION RELATING TO THE ISSUE OF INCORRECT LABEL INFORMATION THROUGH SITUATION ANALYSIS AND CORRECTIVE AND PREVENTIVE ACTIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED PRIOR TO USING PAXGENE® BLOOD DNA TUBE THAT ONE (1) TRAY CONTAINED TUBES WITH TWO DIFFERENT BATCH NUMBERS AND EXPIRATION DATES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
IT WAS REPORTED PRIOR TO USING PAXGENE® BLOOD DNA TUBE THAT ONE (1) TRAY CONTAINED TUBES WITH TWO DIFFERENT BATCH NUMBERS AND EXPIRATION DATES. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 43237 | PAXGENE® BLOOD DNA TUBE | BLOOD/PLASMA COLLECTION DEVICE FOR DNA TESTING | PJE | BECTON, DICKINSON AND COMPANY (BD) | 4095154 | 04053727611659 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |