FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOGUARD BL 22GA X 1.0IN

MDR report key: 20724690 · Received November 19, 2024

Report

Report Number
1710034-2024-01357
Event Type
Malfunction
Date Received
November 19, 2024
Date of Event
October 31, 2024
Report Date
June 16, 2025
Manufacturer
BECTON DICKINSON
Product Code
FOZ
UDI-DI
00382903814237
PMA / PMN Number
K201075
Removal / Correction Number
MDS-25-5274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. ADDITIONAL BATCH REPORTED: BATCH: 4212199, BATCH CREATION DATE: 2024-08-05, BATCH EXPIRATION DATE: 2027-07-31, FULL UDI: (B)(4).

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THE COMPLAINT THAT THE NEEDLE WAS NOT RETRACTING AND THAT THE CATHETER WAS SPLIT COULD NOT BE RETURNED FROM THE UNUSED REPRESENTATIVE 22G INSYTE AUTOGUARD UNITS THAT WERE RECEIVED FROM LOTS 4212199 AND 4229661. A FUNCTIONAL TEST AND MICROSCOPIC EXAMINATION REVEALED NO DAMAGE OR DEFECTS ON THE RETURNED SAMPLES. ALTHOUGH THE REPRESENTATIVE SAMPLES AND MANUFACTURING RECORDS DO NOT SUPPORT THE COMPLAINANT¿S DESCRIPTION OF THE REPORTED EVENT, THE COMPLAINT HAS BEEN DOCUMENTED AND WILL CONTINUE TO BE MONITORED AS PART OF ONGOING EFFORTS TO IDENTIFY POTENTIAL MANUFACTURING RELATED ISSUES. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION OF FA #.

Additional Manufacturer Narrative · 0

CORRECTION OF TYPO IN INVESTIGATION SUMMARY: THE COMPLAINT THAT THE NEEDLE WAS NOT RETRACTING AND THAT THE CATHETER WAS SPLIT COULD NOT BE CONFIRMED FROM THE UNUSED REPRESENTATIVE 22G INSYTE AUTOGUARD UNITS THAT WERE RECEIVED FROM LOTS 4212199 AND 4229661. A FUNCTIONAL TEST AND MICROSCOPIC EXAMINATION REVEALED NO DAMAGE OR DEFECTS ON THE RETURNED SAMPLES. ALTHOUGH THE REPRESENTATIVE SAMPLES AND MANUFACTURING RECORDS DO NOT SUPPORT THE COMPLAINANT¿S DESCRIPTION OF THE REPORTED EVENT, THE COMPLAINT HAS BEEN DOCUMENTED AND WILL CONTINUE TO BE MONITORED AS PART OF ONGOING EFFORTS TO IDENTIFY POTENTIAL MANUFACTURING RELATED ISSUES. A REVIEW OF OUR RISK MANAGEMENT DOCUMENTATION WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY.

Description of Event or Problem · 0

MATERIAL # 381423, BATCH # 4212199, 4229661. IT WAS REPORTED BY CUSTOMER THAT THE NEEDLE NOT RETRACTING AND THE TIP OF THE CATHETER IS SPLIT. VERBATIM: MAT# 381423 LOT# 4212199,4229661 THE NEEDLE NOT RETRACTING AND THE TIP OF THE CATHETER IS SPLIT, CUSTOMER HAS SAMPLES TO SEND. COULD YOU PLEASE PROVIDE A FURTHER DESCRIPTION OF THE ISSUE? REPORTS FROM SEVERAL NURSES STATING THEY INSPECTED THE CATHETER PRIOR TO USE AND THEY LOOK NORMAL, BUT WHEN USED THEY TIP OF THE CATHETER IS SPLIT AND THE PATIENT EXPERIENCES A LOT OF PAIN. WHEN ATTEMPTING TO RETRACT THE NEEDLE, IT WILL NOT RETRACT UNTIL THE BUTTON IS PRESSED SEVERAL TIMES. WAS BOT THE ISSUES "NEEDLE NOT RETRACTING AND THE TIP OF THE CATHETER IS SPLIT" OCCURRED ON BOTH LOT# 4212199 & 4229661? YES, BOTH LOT NUMBERS. WHEN WAS THIS DEFECT OBSERVED? DURING OR BEFORE USE? DURING USE WHAT WAS THE IMPACT TO THE PATIENT? THE PATIENTS HAVE HAD BLOWN VEINS AND COMPLAIN OF A LOT OF PAIN. ALSO, THEY HAVE TO GET STUCK SEVERAL TIMES. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO SERIOUS INJURY, BUT THE DOCTORS HAVE BEEN MADE AWARE OF THE DEFECTS. THEY ARE CONCERNED ABOUT CORRECTIVE ACTION AND COST WHEN WE HAVE TO USE SEVERAL PER PATIENT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1334415 INSYTE AUTOGUARD BL 22GA X 1.0IN CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS FOZ BECTON DICKINSON 4229661 00382903814237

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown