SYNCHRON® LX®I 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-01259
- Event Type
- Malfunction
- Date Received
- April 28, 2011
- Date of Event
- April 2, 2011
- Report Date
- April 2, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023049
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
BCI CUSTOMER TECHNICAL SUPPORT (CTS) HAD CUSTOMER STOP SYSTEM FUNCTION AND PLACE PAPER TOWELS IN COMPARTMENT. CUSTOMER WAS UNABLE TO LOCATE SOURCE OF THE LEAK. CTS RECOMMENDED CUSTOMER TO REPLACE REAGENT SYRINGE AND PRIME REAGENT SUB SYSTEM. CUSTOMER WAS STILL GETTING CUVETTE NOT DRY FLAG. CTS ALSO RECOMMENDED REPLACING WASH TOWER WIPER AND DOING SYSTEM POWER DOWN REBOOT WHICH DID NOT RESOLVE THE PROBLEM. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON-SITE (B)(6) 2011 AND REPLACED THE HAMILTON AND T-VALVES TO STOP REAGENT PROBE LEAKING. FSE ALSO REPLACED LEVEL SENSE BEAD DUE TO WET CUVETTE UPON DISPENSING REAGENT AND REPLACED THE CUVETTE WASH TOWER MANIFOLD.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT LIQUID WAS DRIPPING ONTO COVER FROM REAGENT PROBES AND GETTING CC CUVETTE NOT DRY FLAGS. NO INJURY WAS REPORTED AND NO ERRONEOUS PATIENT RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX®I 725 CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |