FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 2072330 · Received April 28, 2011

Report

Report Number
2050012-2011-01255
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
March 30, 2011
Report Date
March 30, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC PRIOR TO THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES BUT THE LAB'S RANGE IS SET TO 3 SD FOR NA.A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE AND IDENTIFIED SAMPLE TUBE GEL IN THE SAMPLE PROBES OF BOTH INSTRUMENTS. THE FSE REPLACED THE SAMPLE PROBES ON BOTH INSTRUMENTS. THE ROOT CAUSE IS ATTRIBUTED TO POOR SAMPLE QUALITY.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT BOTH THEIR UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS GENERATED 16 FALSE LOW SODIUM (NA) RESULTS WHICH WERE REPORTED OUTSIDE OF THE LABORATORY. WHEN THE PHYSICIANS QUESTIONED THE RESULTS, THE INSTRUMENTS WERE RECALIBRATED AND THE SAMPLES WERE REPEATED. RESULTS WERE HIGHER AND THE REPORTS WERE AMENDED. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN IN THE ATTACHMENT. THE RESULTS CAME FROM BOTH INSTRUMENTS AND THE CUSTOMER CANNOT IDENTIFY WHICH INSTRUMENT GENERATED WHICH RESULTS. THE DETAILS OF THE OTHER INSTRUMENT ARE DOCUMENTED IN MDR #2050012-2011-01265. THERE WAS NO AFFECT TO PATIENT TREATMENT WITH REGARD TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1