FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2072308 · Received April 28, 2011

Report

Report Number
2050012-2011-01263
Event Type
Malfunction
Date Received
April 28, 2011
Date of Event
April 2, 2011
Report Date
April 2, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BCI CTS (CUSTOMER TECHNICAL SUPPORT) HAD CUSTOMER STOP SYSTEM FUNCTION AND PLACE PAPER TOWELS IN COMPARTMENT. THE CUSTOMER WAS UNABLE TO LOCATE SOURCE OF LEAK. A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE (B)(6) 2011 AND FOUND EIC LEAKING FROM THE REAR VALVE. THE EIC WAS REPLACED ABOUT 2 WEEKS AGO. UPON INSPECTING THE EIC WHILE RUNNING, FSE FOUND REAGENT LEAKING FROM THE BACK VALVE DUE TO THE FAULTY VALVE. FSE REPLACED THE WHOLE ASSEMBLY DUE TO RUST IN SCREW PORT OF ACRYLIC BLOCK.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE ELECTROLYTE INJECTION CUP (EIC) WAS LEAKING ONTO THE OVERFLOW PLATE. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1